Feyond-A400 Multi-Mode Microplate Reader, ABS / FL / LUM / TRF
A four-mode microplate reader for labs that run time-resolved fluorescence alongside routine absorbance, fluorescence and luminescence assays. The Feyond-A400 adds TRF with a detection limit of 0.02 pM. Sold in Canada by Luna Nanotech of Markham, Ontario, with prices in Canadian dollars. Contact us with questions or for a quote.
The Feyond-A400 is a four-mode microplate reader that adds time-resolved fluorescence (TRF) to the absorbance, fluorescence and luminescence capabilities of the Feyond-A300. It suits labs running TRF immunoassays alongside routine ELISA, reporter-gene and cell-based assays on a single instrument.
- Four detection modes: absorbance (ABS), fluorescence (FL), luminescence (LUM) and time-resolved fluorescence (TRF).
- Sensitive TRF: detection limit as low as 0.02 pM under optimized conditions.
- Full spectral scanning: any wavelength from 200 to 1000 nm using a monochromator.
- Many plate formats: 6 to 384-well plates, PCR plates, and optional u-Nano ultra-micro plates.
- Incubation and shaking: up to 45 °C, with linear, circular and double-circular shaking.
- Upgradable: optional dual automatic injector and easy-change modular filters.
- Software: ReaderIt-II analysis software.
Catalogue number AS-19060-00.
Technical Specifications
- Model: Feyond-A400 (catalogue no. AS-19060-00), multi-mode: ABS, FL, LUM, TRF
- Plates: 6, 12, 24, 48, 96 and 384-well; 96/384 PCR plates; u-Nano 16/96 ultra-micro plates
- Absorbance: xenon lamp; monochromator 200-1000 nm in 1 nm steps; wavelength accuracy 2 nm, repeatability 0.2 nm, bandwidth under 2.5 nm; range 0-4 OD, resolution 0.0001 OD; accuracy ±(1.0% + 0.003 OD) up to 2.0 OD; linearity R² ≥ 0.999 (0-3 OD); stray light 0.1% at 220 nm; 96-well plate read in under 15 s (fast mode)
- Fluorescence: top reading; xenon lamp; PMT detector; excitation 200-1000 nm, emission 270-850 nm; standard filters EX485/EM530, EX523/EM564, EX624/EM692 (others available); detection limit 1 pM; linear dynamic range 6 logs or more
- Luminescence: PMT detector; detection limit 15 amol/well or better (96-well white plate, flash ATP); crosstalk 0.005% or lower; 200-850 nm; linear dynamic range 6 logs or more
- Time-resolved fluorescence (TRF): filter EX365/EM612; detection limit 0.02 pM (optimized conditions)
- Read types: endpoint, kinetics, spectral scanning, well scanning (up to 900 points per well)
- Shaking: linear, circular, double circular; speed and strength adjustable
- Incubation: room temperature + 4 °C up to 45 °C; ±0.5 °C at 37 °C
- Display: 10-inch capacitive touch screen; 10 GB data storage
- Software: ReaderIt-II PC analysis software (Windows 10 and above, 64-bit); test reports can be uploaded to a PC server by FTP
- Ports: USB A x2, USB B x1, Ethernet x1, RS232 (for injector)
- Optional automatic injector: 1 or 2 channels, 5-1000 µL in 1 µL steps, 125-500 µL/s, accuracy ±1 µL (5-50 µL) or ±2% (51-1000 µL), 25 mL waste collection
- Power: AC 100-240 V, 50-60 Hz
- Dimensions (W x D x H): 420 x 550 x 386 mm
- Weight: 33 kg
Frequently Asked Questions
Which detection modes does the Feyond-A400 have?
Absorbance, fluorescence, luminescence and time-resolved fluorescence (TRF).
How sensitive is the TRF mode?
The detection limit is 0.02 pM under optimized conditions (filter EX365/EM612).
How does the A400 compare with the A300 and A500?
The A300 has absorbance, fluorescence and luminescence. The A400 adds TRF. The A500 adds fluorescence polarization on top.
Which plates can it read?
6, 12, 24, 48, 96 and 384-well plates, 96 and 384 PCR plates, and optional u-Nano ultra-micro plates.